Effects of atmospheric ozone on microarray data quality

Anal Chem. 2003 Sep 1;75(17):4672-5. doi: 10.1021/ac034241b.

Abstract

A data anomaly was observed that affected the uniformity and reproducibility of fluorescent signal across DNA microarrays. Results from experimental sets designed to identify potential causes (from microarray production to array scanning) indicated that the anomaly was linked to a batch process; further work allowed us to localize the effect to the posthybridization array stringency washes. Ozone levels were monitored and highly correlated with the batch effect. Controlled exposures of microarrays to ozone confirmed this factor as the root cause, and we present data that show susceptibility of a class of cyanine dyes (e.g., Cy5, Alexa 647) to ozone levels as low as 5-10 ppb for periods as short as 10-30 s. Other cyanine dyes (e.g., Cy3, Alexa 555) were not significantly affected until higher ozone levels (> 100 ppb). To address this environmental effect, laboratory ozone levels should be kept below 2 ppb (e.g., with filters in HVAC) to achieve high quality microarray data.

MeSH terms

  • Artifacts*
  • Atmosphere / chemistry*
  • Carbocyanines / chemistry*
  • Desiccation
  • Fluorescence
  • Oligonucleotide Array Sequence Analysis / instrumentation*
  • Oligonucleotide Array Sequence Analysis / standards*
  • Ozone / analysis*
  • Ozone / chemistry*
  • Quality Control
  • Reproducibility of Results

Substances

  • Carbocyanines
  • cyanine dye 3
  • cyanine dye 5
  • Ozone